No Products in the Cart
              
            Specifications:
- Height: 9.2 In (23.8 cm)
- Width: 4.5 In (11.3 cm)
- Depth: 2 In (5 cm)
- Weight: 2 lbs (1 kg)
- Syringes: Monoject or B-D 60 cc, Monoject or B-D 20 cc
- Power:
  - Operating Voltage: 5-7 Volts
  - Operating Current: 20mA
  - Low Battery Voltage: 5.0 Volts
- Battery:
  - Type: Alkaline (NEDA 14A)
  - Quantity: 4 (C Size)
  - Battery Life (New): 150 Hours
  - Battery Life (At Low Battery): 4 Hours
- Accuracy:
  - Infusion: Linear Rate ±3%
  - Bolus: Linear Displacement ±3%
- Flow:
  - Flow Rates: 0 to 600 mL/hr, depending on the Smart Label
  - Flow Profile: Bolus - Continuous, Infusion - Pulsed Intermittently
- Force and Pressure:
  - Occlusion Force: 8 ± 1 lbs. (pressure)
  - Max Occlusion Pressure: 11 psi with 60 cc syringe, 20 psi with 60 cc syringe
  - Occlusion Detection Time: 120 seconds at 36 mL/hr
  - Volume Stored on Occlusion: 1.1 mL Approximate
          
