Specifications:
- Height: 9.2 In (23.8 cm)
- Width: 4.5 In (11.3 cm)
- Depth: 2 In (5 cm)
- Weight: 2 lbs (1 kg)
- Syringes: Monoject or B-D 60 cc, Monoject or B-D 20 cc
- Power:
- Operating Voltage: 5-7 Volts
- Operating Current: 20mA
- Low Battery Voltage: 5.0 Volts
- Battery:
- Type: Alkaline (NEDA 14A)
- Quantity: 4 (C Size)
- Battery Life (New): 150 Hours
- Battery Life (At Low Battery): 4 Hours
- Accuracy:
- Infusion: Linear Rate ±3%
- Bolus: Linear Displacement ±3%
- Flow:
- Flow Rates: 0 to 600 mL/hr, depending on the Smart Label
- Flow Profile: Bolus - Continuous, Infusion - Pulsed Intermittently
- Force and Pressure:
- Occlusion Force: 8 ± 1 lbs. (pressure)
- Max Occlusion Pressure: 11 psi with 60 cc syringe, 20 psi with 60 cc syringe
- Occlusion Detection Time: 120 seconds at 36 mL/hr
- Volume Stored on Occlusion: 1.1 mL Approximate