No Products in the Cart
Refurbished Philips M1034A BIS Module now available. This module is employed to measure the Bispectral Index (BIS). The BIS System comprises the Digital Signal Converter (DSC), BIS Engine, and BIS Module. To conduct a unit test, connect the Bis Sensor Simulator while the monitor is in Monitoring Mode. Enter the Bis Menu and select Show SENSOR, then initiate the Impedance check by pressing Start Cyclic Check. The expected results are as follows:
- Electrode 1 (+): 4-6 kOhms
- Electrode 2 (ref): 8-17 kOhms
- Electrode 3 (1-): 2-4 kOhms
- Electrode 4 (2-): 3-5 kOhms
Additional OEM Part Numbers include M1034-60201 and 453563461411.