No Products in the Cart
Modular patient monitoring system is convenient, cost-effective
The Mindray V12 is a versatile patient monitor that features a modular design fully scalable to meet the demands of a variety of clinical applications. Designed to serve as both a bedside and transport monitors, medical staff can simply remove the portable V12 from the VDock to ready the unit for transport — without needed to disconnect cables or move modules. Featuring color-coded patient monitor connectors with unique key-code feature to help clinicians quickly respond to a patient’s changing condition.
The Mindray V12 Patient Monitor is available professionally refurbished from Avante Health Solutions, along with a full suite of related parts and services. For more information, contact an Avante representative today.
Features
Monitoring parameters include: ECG, NIBP, HR, IBP, TEMP, SpO2, CO2.
Versatile patient monitor with integrated module bay.
Designed to serve as both a bedside and transport monitor.
Modular design is fully scalable to meet the needs of each care unit.
Interchangeable system elements for added convenience, affordability.
Integrated wireless capability.
Color-coded patient monitor connectors.
Unique, key-code feature.
For full availability information, contact an Avante representative.
SPECIFICATIONS
Display
Type: 12.1 in color TFT touchscreen display
Resolution: 1024 x 768
Aspect ratio: 3:4 in portrait mode
Waveforms: 8 selectable
Physical Dimensions
Height: 13.7 in (34.8 cm)
Width: 9.4 in (23.9 cm)
Depth: 7.5 in (19.1 cm)
Weight: 8.9 lb (4 kg)
Weight (with module and 3 batteries): 13 lb (5.9 kg)
Heart Range Meter
Measurement Range - Adult/Pediatric: 30 – 300 bpm
Measurement Range - Neonate: 30 – 350 bpm
Resolution: 1 bpm
Accuracy: 30 – 250 bpm: ±3 bpm or ±3%, whichever is greater 251 – 350 bpm: ±5%
ST Segment Analysis
Adult/Pediatric Only
Measurement Range: –10.0 mm to +10.0 mm (-1000?V to +1000?V RTI)
Resolution: 0.1 mm (10?V)
User Selectable ST Measurement Points: 40, 60 or 80 ms after the J point (HR independent); 60/80 HR dependent
Respiration (ECG)
Measurement Range: 4 – 199 rpm
Resolution: 1 rpm
Accuracy: 4 – 150 rpm: ±2 rpm or ±2%, whichever is greater 155 – 199 rpm: ±4%
Sensing Lead: II or I (default: lead II)
Sweep Speed: 3.125, 6.25, 12.5, 25 mm/sec
Respiration (CO2)
Measurement Range: 0 – 150 rpm
Resolution: 1 rpm
Accuracy: 0 – 70 rpm: ±1 rpm | 71 – 120 rpm: ±2 bpm | 121 – 150 rpm: ±3 bpm
Non-Invasive Blood Pressure
Measurement Method: Oscillometric
Measurement Modes: Manual, Interval, Stat
Static Pressure Range: 0 – 300 mmHg
Accuracy: ± 3 mmHg
Invasive Blood Pressure
Resolution: 1 mmHg
Accuracy: ±2 mmHg or 2% of the signal value, whichever is greater (excluding transducer)
Zero Offset Range: ±120 mmHg
Zero Accuracy: ±1 mmHg
Excitation: 5 Volts DC, ±2%
Frequency Response: 8 Hz ±1 Hz , 12 Hz ±1 Hz, 16 Hz ±1 Hz, (-3 db Point)
Pulse Oximetry - Masimo SET SpO2
Resolution: 1%
Accuracy: Adult: ±2% with no motion | ±3% during motion conditions, with motion resistant sensors
Pediatric: ±2% with no motion | ±3% during motion conditions, with motion resistant sensors
Neonate: ±3% with no motion | ±3% during motion conditions, with motion resistant sensors
Pulse Oximetry - Nellcor OxiMax SpO2
Resolution: 1%
Accuracy: 70–100% ±2% for sensors: MAX-A, MAX-AL, MAX–P, MAX-I, MAX-N (Adult), MAX-N (Neonate) | 70–100% ±2.5% for sensors: OxiCliq A, OxiCliq P, OxiCliq I, OxiCliq N (Adult), OxiCliq N (Neonate) | 70–100% ±3% for sensors: DS-100A, D-YS ( Infant to Adult), OXI-A/N (Adult), OXI-P/I | 70–100% ±3.5% for sensors: D-YS with D-YSE Ear Clip, D-YS with D-YSPD Spot Clip | 70–100% ±4% for sensors: D-YS (Neonate), OXI-A/N (Neonate)
Temperature
Unit of Measure: Selectable °C or °F
Measurement Range: 15 – 45°C, inclusive | 59 – 113°F, inclusive
Resolution: 0.1°C, 0.1°F
Accuracy: ±0.1°C (15 – 45°C), exclusive of probe errors | ±0.2°F (59 – 113°F), exclusive of probe errors
CO2 Microstream® Module
Measurement Range: 0 – 99 mmHg, inclusive (at sea level)
Accuracy: 0 – 38 mmHg: ±2 mmHg | 39 – 99 mmHg: ±(5% of reading + 0.08% for every 1 mmHg above 38 mmHg)
*Accuracy applies for breath rates up to 80 bpm. For breath rates above 80 bpm, accuracy is 4 mmHg or ±12% of reading, whichever is greater, for EtCO2 values exceeding 18 mmHg.
Flow Rate: 50 ml/min -7.5 +15 ml/min
Environmental Conditions
Storage/Transport Temperature: -20 – 60°C (-4 – 140°F)
Storage/Transport Humidity: 10 – 95%, non-condensing
Storage/Transport Altitude: 1060 – 570 hPa
Operating Temperature: 5 – 40°C (41 – 104°F)
Operating Humidity: 15 – 95%, non-condensing
Operating Altitude: 1060 – 700 hPa
Electrical Ratings
Battery Type: Lithium Ion
Number of Batteries: 3
Battery Voltage: 11.1 V
Battery Capacity: 4800 mAh
Battery Runtime: 4 hours with three fully charged batteries while monitoring 6-lead ECG, Respiration, Masimo SpO2, 2 IBPs, CO2, Temperature and 1 NIBP and print strip every 15 minutes
Battery Recharge Time: 4 hrs maximum