B-Braun Outlook 400es

    B-Braun Outlook 400es

    $0.00
    DESCRIPTION

    Dimensions:

    - Height: 4.5 inches (11.43 cm)
    - Width: 12 inches (30.48 cm)
    - Depth: 8.75 inches (22.23 cm)
    - Weight: 11 pounds (4.98 kg)

    Infusion System:

    - Delivery Rate Range: 0.1 to 999.9 ml/hr, in 0.1 ml/hr increments; Micro Mode: 0.1 to 99.9 ml/hr
    - Volume to be Delivered Range: 0.1 to 9999.9 ml, in 0.1 ml increments
    - Occlusion Limit Settings: 75, 100, 200, 300, 400, 500, and 750 mmHg. Standard default is 300 mmHg.
    - Keep Vein Open Rate: 3.0 ml/hr or selected rate, whichever is lower
    - Volume Delivered Accuracy: + 5%
    - Fluid Types: All standard IV fluids

    Alarms:

    - Alarms: Air-In-Line, Battery Very Low, Close Roller Clamp, Container Empty, Downstream Occlusion, Door Open, Hold Time Exceeded, KVO, Low Battery Warning, System Error.
    - Air-in-Line Alarm: > 100 microliter bubble
    - Memory: Permanent data retention after the pump is turned off unless the battery is depleted or disconnected, or unless the Data Retention Defaults have been otherwise selected in the Biomed Options.

    Max Bolus Volume (Microbore Tubing):

    - 0.047 ml at 399 ml/hr (75 mmHg)
    - 0.086 ml at 399 ml/hr (750 mmHg)
    - 0.053 ml at 500 ml/hr (400 mmHg)
    - 0.097 ml at 750 ml/hr (750 mmHg)