Imed Gemini PC-2TX infusion Pump

    Imed Gemini PC-2TX infusion Pump

    $0.00
    DESCRIPTION

    Product Specifications
    Operating Principle:
    Linear peristalsis
    Dimensions:
    10.6? x 11.9? x 7.2? (26.9 cm x 30.2 cm x 18.3 cm)
    Weight:
    18.5 lbs (8.4 kg)
    Power Requirements:
    90 – 130 (120 nominal) VAC, 0.4A, 50/60 Hz, 3 Wire, single phase
    Battery Operation:
    With a new, fully charged battery, approximately 5 hours with one channel operating at 125 ml/hr, or approximately 4 hours with two channels operating at 125 ml/hr.
    Electrical Leakage:
    Less than 100 microamps
    Operating Range
    Controller Mode:
    0.1-500 ml/hr
    Pump Mode:
    0.1-999 ml/hr
    KVO (Keep Vein Open) rate:
    1 ml/hr
    Volume to be Infused range:
    1-9999 ml
    Occlusion Pressure
    Pump Mode:
    10 psi for rates above 30 ml/hr.For rates Below 30 ml/hr, the occlusion pressure is Rate dependent.
    Advisories:
    Secondary, Primary, Infusion Complete – KVO, Empty Container – KVO, Low Battery, Maintenance, Pause, Low Flow, Total Vol Infused, and Sec Vol Infused.
    Prompts:
    Select PC Mode, Set Rate, Set VTBI, Press Start, and Access Channel
    Alarms:
    Check ECD, Check IV Set, Close Door, Air-in-line, Occluded Patient Side, Occluded Fluid Side, Occluded, Partial Occlusion Fluid Side, Flow-Stop Open/Close Door, KVO Comm Tout, and KVO
    For service, parts, PM or rental see our Services section