Philips M1012A C.O. Module

    Philips M1012A C.O. Module

    $0.00
    DESCRIPTION

    The Phillips M1012A Cardiac Output Module provides measurements of Cardiac Output (CO) and Continuous Cardiac Output (CCO). Two measurement methods are available: the PiCCOTM 1 method, for assessing CO, COO, and other hemodynamic parameters, and the Right Heart Thermodilution method for measuring CO. Both methods are suitable for adult, pediatric, and neonatal patients in critical care settings.

    This module can be seamlessly connected through plug-and-play to any monitor in the V24 (M1205A), CMS (M1165A/66A/75A and M1175A/76A/77A), and IntelliVue (M8005A, M8007A, M8010A) families. The module transmits a thermodilution curve, cardiac output numerics, and cardiac index values to the patient monitor display.

    Continuous Cardiac Output (COO) Features:
    - COO can be measured using the Transpulmonary Thermodilution or PiCCO Method, developed by Philips in collaboration with Pulsion Medical Systems.
    - Values for Intrathoracic Blood Volume (ITBV) and Extravascular Lung Water (EVLW) can be derived from the Transpulmonary Thermodilution CO measurement (excluding the USA).
    - COO alarms can be generated from Pulse Contour Analysis when the arterial catheter used for Transpulmonary Thermodilution is connected to the module.
    - Continuous Systemic Vascular Resistance values can be derived from the COO.

    Cardiac Output (CO) Features:
    - Absolute and indexed hemodynamic values are displayed.
    - Alarm generation from blood temperature measurement.
    - Patient isolation ensured by internal optical coupling.
    - Hard keys on the module for direct access to the most commonly performed functions.
    - Light indicators identify the module in use during monitoring tasks.
    - Color-coded patient cable connectors for quick and easy identification.
    - Compact size due to surface mount technology.

    Cardiac Output Module Includes:
    - Up to six cardiac output curves and values displayed simultaneously.
    - Seven curve alerts.
    - Easily selectable or rejectable cardiac output curves.
    - Injectate temperature.
    - Computation constant.
    - Averages of selected curves made with further selectors or rejectors.

    Patient Monitor Cardiac Output Module Specifications:
    - Dimensions: Height: 3.9 in (99.6 mm), Width: 1.4 in (36 mm), Depth: 3.8 in (97.5 mm), Weight: 225 g (7.9 oz.)
    - Temperature Range: Operating: 0 to 55ºC (32 to 131ºF), Storage: -40 to 70ºC (-40 to 158ºF)
    - Altitude Range: Operating: Up to 4,600m (15,000ft), Storage: Up to 15,300m (50,000ft)
    - Compatibility: If used in a monitor without PiCCO functionality, only the right heart method can be employed (e.g., CMS monitor release earlier than C.0 or a V24/V26 monitor).

    ADDITIONAL INFORMATION
    REVIEWS

    RECENTLY VIEWED PRODUCTS

    BACK TO TOP