ValleyLab Force FXC ESU Monitor

    ValleyLab Force FXC ESU Monitor

    $0.00
    DESCRIPTION

    Instant Response™ Technology
    Instant Response™ technology provides surgeons with improved performance at lower power settings, minimizing the risk of tissue damage and neuromuscular stimulation, and decreasing the need to "turn up the generator."

    A Smoother Cut Through All Tissue Types
    The Force FX™ generator adjusts automatically, responding to tissue changes, maintaining power delivery, and minimizing drag.

    Added Safety and Reliability Minimizes Electrosurgical Risks
    Capacitive coupling is reduced by 30-50% when using Instant Response™ technology. This reduction is achieved by limiting the RMS voltage and the high frequency harmonics. Lower voltage means less neuromuscular stimulation and more precise delivery of energy to reduce collateral damage.

    What is Instant Response™ Technology?
    Instant Response™ technology features an advanced feedback system that recognizes changes in tissue 200 times per second, and adjusts voltage and current accordingly to maintain appropriate power. This unique capability differentiates Instant Response™ generators from any other, as shown by their high power efficiency rating (PER). That's smart™ generator technology.

    PER indicates the ability of a generator to accurately deliver the selected power into a wide range of tissue types. With a PER of 98, the Force FX™ generator provides a consistent surgical effect and has a higher PER than any other electrosurgical generator.

    Instant Response™ Technology provides consistent power through different tissue types.
    The Force FX™ generator provides surgeons with the advantage of accurately delivering the selected power setting, even through resistant tissue.

    When a conventional generator encounters tissue changes, both power delivery and cutting efficiency are reduced. This means that higher power settings are required in order to cut through all types of tissue. With higher power settings, more energy is delivered, and tissue damage increases.

    Features
    Instant Response™ technology ensures that the power delivered remains virtually constant, regardless of the tissue type
    Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
    Three internal microcontrollers reduce system reaction time and increase the system's processing speed
    Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
    A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
    Three cut modes, all controlled by Instant Response™ technology, offer surgeons a variety of choices:
    Low Cut for delicate tissue or laparoscopic cases
    Pure Cut for a clean, precise cut
    Blend for cutting with hemostasis
    Four coag modes:
    Desiccate for low voltage contact coagulation suitable in laparoscopic and delicate tissue work
    Fulgurate (high crest factor) for efficient noncontact coagulation in most applications
    Fulgurate (low crest factor) for lower voltage coagulation requirements
    Spray for coagulating large tissue areas with superficial depth of necrosis
    Three bipolar modes:
    Precise, Standard, and Macrobipolar are controlled by the Instant Response™ system
    Precise and Standard setting utilize low voltage to prevent sparking
    Versatile system that is uniquely compatible with other devices, including:
    Force Argon™ II and Force GSU™ argon coagulation system
    CUSA EXcel™ and CUSA™ 200 ultrasonic surgical aspirators
    OptiMumm™ smoke evacuator, through a direct cable link
    Valleylab VLCM bipolar current monitor
    Compatible with and used as the electrosurgical energy source for:
    Dyonics* Control RF arthroscopic ablation system
    Dyonics* Electroblade™ rotary resection system
    Cook Vascular Perfecta™ EDS pacemaker lead extraction system
    Compatible with, and the exclusive electrosurgical generator for, the Computer Motion Hermes™ Voice Command System
    * Trademark of Smith & Nephew

    SPECIFICATIONS
    Output Waveforms
    Bipolar
    Precise: 470 kHz sinusoid
    Standard: 470 kHz sinusoid
    Macro: 470 kHz sinusoid
    Monopolar Cut
    Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
    Pure: 390 kHz sinusoid
    Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
    Monopolar Coag
    Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
    Fulgurate: 390 kHz damped sinusoid bursts with a repetition frequency of 30 or 57 kHz into 500 ohms.
    Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21 kHzz < f < 35 kHz. Output is further modulated by a random 250 Hz envelope with a variable duty cycle.
    Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208-264 volts (at rated load).

    Low Frequency Leakage (50-60 Hz)
    Source current, patient leads, all outputs tied together.

    Normal polarity, intact chassis ground: < 10 µA
    Normal polarity, ground open: < 50 µA
    Reverse polarity, ground open: < 50 µA
    Sink current, 140V applied, all inputs: < 50 µA
    PER = 98
    PER (Power Efficiency Rating) is the measure of an electrosurgical generator's ability to accurately deliver the selected power into a wide range of tissue types.

    Physical Weight & Dimensions
    Height: 57 inches
    Width: 51-1/2 inches
    Depth: 24-1/8 inches
    Input Power Requirements
    Operating range is 85 to 132 AC volts. Maximum current is 7 amperes in Cut and 4 amperes in Coag.

    High Frequency Leakage
    Bipolar: Less than 60 mA (rms)
    Monopolar: Less than 150 mA (rms)
    REM™ Contact Quality Monitoring System
    Measurement Frequency: 80 kHz ± 10 kHz
    Measurement Current: Less than 10 µA
    Acceptable Resistance Ranges:
    REM™ pad – 5-135 ohms
    Non-REM™ pad – less than 20 ohms
    Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.
    Adaptive REM™: REM™ trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.
    Output Characteristics
    Bipolar
    Precise Mode:
    Maximum P-P Voltage: 450
    Rated Load (OHMS): 100
    Maximum Power (Watts): 70
    Crest Factor* (Typical): 1.5
    Standard Mode:
    Maximum P-P Voltage: 320
    Rated Load (OHMS): 100
    Maximum Power (Watts): 70
    Crest Factor* (Typical): 1.5
    Macro Mode:
    Maximum P-P Voltage: 750
    Rated Load (OHMS): 100
    Maximum Power (Watts): 70
    Crest Factor* (Typical): 1.5
    Monopolar Cut
    Low Mode:
    Maximum P-P Voltage: 1350
    Rated Load (OHMS): 300
    Maximum Power (Watts): 300
    Crest Factor* (Typical): 1.5
    Pure Mode:
    Maximum P-P Voltage: 2300
    Rated Load (OHMS): 300
    Maximum Power (Watts): 300
    Crest Factor* (Typical): 1.5
    Blend Mode:
    Maximum P-P Voltage: 3300
    Rated Load (OHMS): 300
    Maximum Power (Watts): 200
    Crest Factor* (Typical): 2.5
    Monopolar Coag
    Desiccate Mode:
    Maximum P-P Voltage: 3500
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 5
    Fulg. High Crest Factor* Mode:
    Maximum P-P Voltage: 8500
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 7.0
    Fulg. Low Crest Factor Mode:
    Maximum P-P Voltage: 6900
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 5.5
    Spray Mode:
    Maximum P-P Voltage: 9000
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 8
    Monopolar Coag
    Desiccate Mode:
    Maximum P-P Voltage: 3500
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 5
    Fulg. High Crest Factor* Mode:
    Maximum P-P Voltage: 8500
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 7.0
    Fulg. Low Crest Factor Mode:
    Maximum P-P Voltage: 6900
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 5.5
    Spray Mode:
    Maximum P-P Voltage: 9000
    Rated Load (OHMS): 500
    Maximum Power (Watts): 120
    Crest Factor* (Typical): 8
    CEM™ Mode
    Monopolar Cut (Low):
    Maximum P-P Voltage: 1000
    Rated Load (OHMS): 300
    Maximum Power (Watts): 100
    Crest Factor* (Typical): 1.5
    Monopolar Coag (Dessicate):
    Maximum P-P Voltage: 3500
    Rated Load (OHMS): 500
    Maximum Power (Watts): 70
    Crest Factor* (Typical): 5
    * Crest Factor is an indicator of a waveform's ability to coagulate without cutting.